| Name: |
Human IL-18(Interleukin 18) ELISA Kit |
|||
| Adapted species: | Human | |||
| Specifications: | 48T/96T | |||
| Storage: | Store at 2-8 ℃ for 6 months | |||
| Application: | Human IL-18(Interleukin 18) ELISA Kit can quantitatively determine the concentration of IL-12 in plasma, tissue homogenates, and other biological fluids in vitro | |||
| Testing method: | Double Antibody Sandwich Method | |||
| Detection range: | 15.625-1000pg/ml | |||
| Sensitivity: | < 9.375pg/ml | |||
| Standard curve: | ![]() |
|||
| Rate of recovery: | Add a certain amount of IL-18 to the samples listed in the table below, and calculate the recovery rate by comparing the measured values with the expected amount of IL-18 in the samples | |||
| sample | Recycling range (%) | Average value (%) | ||
| Serum (n=5) | 95-101 | 99 | ||
| EDTA plasma (n=5) | 87-96 | 91 | ||
| Heparin plasma (n=5) | 92-103 | 98 | ||
| Linearity: | The linearity of the reagent kit is determined by detecting samples and their series of dilutions mixed with appropriate concentrations of IL-18, expressed as a percentage of the detected concentration relative to the expected concentration | |||
| sample | 1:2 | 1:4 | 1:8 | |
| Serum (n=5) | 94-103% | 88-105% | 85-97% | |
| EDTA plasma (n=5) | 82-99% | 83-101% | 88-96% | |
| Heparin plasma (n=5) | 80-100% | 87-96% | 85-100% | |
| Coefficient of variation (%): | Intra batch difference: CV<8% Inter batch difference: CV<10% | |||
| Matters needing attention: | For scientific research purposes only | |||